van Aken, J. M., "Evaluation of calibration results from three facilities for a multi-component, multi-piece balance," Proceedings of the 44th International Instrumentation Symposium, pp. 160-179, Reno, Nevada, May 3-7, 1998.

Keywords: Multi-component balance, multi-piece balance, balance calibration

Abstract: A 1.5 inch diameter, multi-piece, six-component (five force / one moment) balance was calibrated at three balance calibration facilities. The main purpose for this exercise was to determine the repeatability of the balance calibration process and resulting calibration matrix accuracy in estimating the applied balance calibration loading at each facility and between the considered facilities.


| Technical Publications | 1998 Publications | Website Menu | Search |